Product Summary |
|
|
Fixed Temp. Incubator from small to large capacity, high functional and best-selling for long years |
|
|
|
|
Features |
|
|
Compact model ideal for sample preservation and constant temperature testing. |
|
The Peltier element (semiconductor element) is used for the cooler, realizing non-fluorocarbon and low vibration. |
|
The temperature controller with program is used. |
|
Option to mount on an air jacket and stack units. |
|
The observation window of the IJ102W is an environmentally friendly combination of energy-saving double-glazing and semi-strengthened glass, which prevents unnecessary heat dissipation. |
|
Two-stage stacking is possible. (Use the optional two-stage stacking rack for IJ101/101W/300 and stacking bracket for IJ201) |
|
Enhanced safety with the addition of self-diagnostic functions, auto overheat prevention, overheat prevention device and over current circuit breaker. |
|
|
|
|
|
|
|
Specifications |
|
|
Specification Item |
Details |
|
Model |
IJ202 |
|
Type |
Forced convection |
|
Operating Temp. Range |
5¡«60¡æ£¨at room temp.25¡æ£© |
|
Temp. Control Accuracy |
±0.5¡æ£¨at 37¡æ£© |
|
Temp. Distribution Accuracy |
±1.0¡æ£¨at 37¡æ£© |
|
Time to Max. Temp. |
20¡«60¡æ¡¡50min. |
|
Time to Lowest Temp. |
20¡«5¡æ¡¡120min. |
|
Interior |
Stainless steel |
|
Heater |
Pipe heater 150W |
|
Condenser |
Peltiert element, Forced radiation type |
|
Temp. Control Method |
PID |
|
Operation Function |
Program operation (30 steps x 1, 15 steps x 2, 10 steps x 3), Fixed operation, Quick auto-stop, Auto-stop, Auto-start operation |
|
Additional Functions |
Calibration off-set, Key-lock function, Auto recovery function at power failure |
|
Sensor |
Double sensor: Platinum resistance Pt 100Ω (Temp. controller), K-thermocouple(Overheat preventor) |
|
Safety Functions |
Self-diagnostic function(Temp. sensor error, heater failure, SSR short-circuit, Auto overheat), Overheat prevention(IJ201/301), Liquid pressure overheat preventor, Over-current ELB, Main relay error |
|
Shelf Loading Capacity |
Approx. 10kg/ shelf |
|
Shelf Step |
8 steps |
|
Accessories |
Shelf(Stainless punching metal) 2pcs., Shelf support 4 pcs., Receiver for drain 1pc. |
|
Temp. Setting Method |
Dedicated operation menu key and digital setting by UP/DOWN key |
|
Internal dimensions£¨W×D×H mm£© |
300×300×300mm |
|
External dimensions£¨W×D×H mm£© |
580×417×437mm |
|
Internal capacity |
27L |
|
Standard Power Source |
AC100V 4.5A |
|
Power Source (Modification)* |
N/A (External transformer required) |
|
Weight |
25kg |