Water Bath (High Precision Constant Temp., Programmable) (BH501)
Bench top precision constant temperature bath equipped with high performance controller for higher precision and wider temperature range.
Features of Water Bath (High Precision Constant Temp., Programmable) (BH501)
Features
Precision controller enables temperature adjustment accuracy of ±0.01°C (at 20°C).
Features
BH501 can be used as precision water or oil bath.
Features
Maximum of 99 steps of programming operation, fixed temperature operation, Quick auto stop, Auto stop and Auto start functions.
Features
RS485 communication function, Temperature output terminal (1~5V), Alarm output terminal, calibration offset function and key lock function.
Features
Better external circulation with powerful circulation pump.
Features
Circulation pump flow changeable up to 10 patterns.
Features
Circulated water temperature (closed type) can be controlled precisely by external sensor.
Features
Control panel can be removed for remote control by using communication cable (sold separately).
Safety features
Self diagnostic functions, Earth leakage circuit breaker, Empty boiling prevention switch, Automatic overheating prevention, Independent overheating prevention.
Specifications
Specifications
Item Code | 222058 |
---|---|
Model | BH501 |
Pump max. flow£¨50/60Hz£© | 14/15L/min. (Circulation pressure loss when below 10kPa) |
Pump max. lift£¨50/60Hz£© | 2.8/3.4m (water flow about 0.5L/min) |
Circulation system | Circulation system Water jet pump circulation (closed system circulation) |
Operating temperature range | RT+15~200°C |
Temperature adjustment accuracy | ±0.01°C (water: RT+15°C~80°C), ± 0.1 (silicon oil: KF96/50cst 70°C~200°C) |
Temperature indication unit | 0.1¡æ |
Circulation Pump | Water jet pump circulation |
Heater | 1.2kW |
Ambient temperature range | 5¡«30¡æ |
Temperature control / system | PID control by microprocessor |
Sensor | W sensor: Pt 100Ω A-class (for temperature control)+ K-thermocouple (for overheat prevention) double sensor |
Timer | 1min.~99hrs.59min., 100~9999hrs Timer / Time switchable function |
Operation function | Fixed temp. operation, quick auto stop, auto stop / start at setting time, program operation (max. 99 steps, repeating, gradient operation), program auto start operation |
Safety device | Earth leakage circuit breaker, key lock function, breakout protection function, automatic overheating prevention, independent overheating prevention, float switch |
Other functions | RS485 communication function, Temperature output terminal, alarm output terminal, calibration offset function |
Accessories | Hose nipple (1/2×ø14mm) 2pcs. |
Internal dimensions£¨W×D×H mm£© | 239×299×200mm |
External dimensions£¨W×D×H mm£© | 310×396×607mm |
Remarks(dimensions) | Effective tank dimensions W227×D150×H200mm£¨Frontage corner R34£©¡¡Tank height 307©L£¨rubber height included£©Internal capacity 13L£¨Liquid volume approx. 10L£© |
Power Source |
AC220V 6A Single-phase with step down transformer |
Weight | 20kg |
Maintenance and warranty
Conditions: temperature and humidity: 23 deg.C+/-5 deg.C, 65%RH+/-20% (no load)
External dimensions do not include protrusions.