Natural Convection Oven (Programmable) (DVS602)
Highly practical standard and programmable ovens
Features of Natural Convection Oven (Programmable) (DVS602)
Programmable natural convection, constant temperature oven with quickly performed program settings
Operation and functions
- Excellent temperature accuracy
- Easy to use and maintain
- Equipped with a 6 pattern PID program controller with easy program settings (30 steps x 1, 15 steps x 2, 10 steps x 3)
- Simultaneously display of set constant and measured temperature
- Quick Auto stop, Auto Start / Stop operation
- Increased safety and self-diagnostic function
-
With calibration off-set function
Safety features
Self diagnosis functions (Temp. sensor abnormal, Heater disconnection, Internal communication error, temperature input circuit abnormality, Automatic overheat prevention function, SSR-short), Overheat prevention, Electric leakage breaker with over current protection
Specifications
Specifications
Item Code | 212863 |
---|---|
Model | DVS602 |
Method | Natural gravity convection |
Operating temp. range | Room temp.£«5¡«260¡æ |
Temp. control accuracy | ±1¡æ£¨at 260¡æ£© |
Temp. distribution accuracy | ±5¡æ£¨at 260¡æ£© |
Max. temp. reaching time | Approx. 90 min. (Room temp. +5¡æ to 260¡æ) |
Interior material | Stainless steel |
Exterior material | Cold rolled steel plate with melamine resin baking finish |
Heat insulating material | Glass wool |
Heater | Stainless pipe heater |
Heater | 1.36kW |
Observation window | 250×280 mm Chemically strengthened glass×3 |
Cable hole | 30 mm l.D.×1 pcs.(right side) |
Exhaust port | 30 mm l.D.×2 pcs.(on top) |
Temp. controller | 3 patterns program controller, PID control by microprocessor |
Temp. setting method | Operation menu key and digital setting by UP/DOWN key |
Temp display method | Measurement temp. : Digital display by green LED Setting temp. : Digital display by red LED |
Timer | 1 min. to 99 hrs. 59 min. and 100 hrs. to 999 hrs. 50 min. |
Operation function | Fixed temperature operation, Program operation, Auto start, Auto stop, Quick Auto-stop |
Program mode | Program operation: 6 pattern, 30 steps (30 steps×1, 15 steps×2, 10 steps×3) |
Additional functions | Calibration off-set function, Key lock, Uninterruptible power for memory, Pattern repeat function |
Heater circuit control | SSR control |
Sensor | K-thermocouple |
Safety device | Self diagnosis functions (Temp. sensor abnormal, Heater disconnection, SSR- short, Memory abnormal |
Safety device | Automatic overheating prevention), Overheat prevention, Electric leakage breaker with over current protection |
Shelf plate load | Approx. 15kg / pcs. |
Shelf support step number / pitch | 13 steps / 30mm |
Shelf plate / Shelf bracket | Stainless steel, 2 pcs. / 4 pcs. |
Internal dimensions£¨W×D×H mm£© | 600×540×500mm |
External dimensions£¨W×D×H mm£© | 710×651×870mm |
Remarks(dimensions) | 162L |
Power Source |
AC220V 7A Single-phase |
Weight | 63kg |
Maintenance and warranty
Conditions: temperature and humidity: 23 deg.C+/-5 deg.C, 65%RH+/-20% (no load)
External dimensions do not include protrusions.