Extraction Oven System OV-SCF-10000 Series
Extraction Oven System OV-SCF-10000 Series Micro-Research System SCF CO2 Extraction Equipment Extraction Machine Selangor, Malaysia, Kuala Lumpur (KL), Puchong Supplier, Distributor, Supply, Supplies | Renetech Sdn Bhd
Oven Extraction Furnace
Specifications
1. Operation temperature: room temperature ~220 ¡æ
2. Operation pressure: Operating pressure up to 7250 psi (MWAP=7250 PSI)
3. Operation capacity: Optional 60 / 250 / 1000ml High-pressure extraction vessels
4. Double extraction loop: Optional 1 or 2 extraction loop
5. Extraction mode: dynamic and soak extraction
6. Cleaning method: capable of solvent injection device
7. Flow control: Adjustable flow rate control with display of CO2 flow rate
8. Certification: Whole system pass CE
        IEC 61010-2-010 : 2002 (2nd Ed) for use with IEC 61010-1 : 2001
        EN 61010-2-010 : 2004 (2nd Ed) for use with EN 61010-1 : 2002 Standard
9. Dimension: 700 mm × 650 mm × 950 mm
10.Weight: 80 kg
 
Utility Needs
1. Power supply: 110V , 16A
 
Application Focus
1. Supercritical , liquid CO2 extraction
2. Provide column chromatography or separation column for hot air heating
 
---------------------------------------------------------------------------------------------------------------
Digital Pressure Control Module(PM-10000C)
Specifications
1. System output pressure: 1500~10000 psi
    (depending on driving pressure)
2. System output flow rate: 200ml/min (up to 4350 psi)
3. Booster pump
4. Built-in pre-cooler (without chiller)
5. Piping: seamless stainless steel tube
6. Certification: Whole system pass CE
    IEC61010-2-010(2nd Ed) for use with IEC61010-1:2001  Standard
7. Dimensions: 360 mm × 600 mm × 270 mm
8. Weight: 30 kg
 
Control System
1.Touch-Panel for Pressure Control Systems and human-machine operation interface
2. Programmable pressure control system
3. Proportional precise pressure control
4. Multi-step for booster and immediately adjust / response control
5. Alarm (notice) for abnormal operation and operation timer
 
Safety Protection
1. Overpressure abnormal alarm devices
2. Emergency stop devices and automatic gas cut-off installations for overpressure
3. Safety relief valve for overpressure (full flow rate)
 
Utility Needs
1. Power supply: 110/220V
2. Source air: (5~8 kg/c©O)
3. Chilly: water temperature≤5¡æ
 
Application Focus
1. Liquid CO2 transmission and booster
2. Supercritical CO2 cleaning equipment applications

Please leave your enquiry here, we will reply as soon as possible.


*only support gif, jpeg, jpg, png, pdf
Switch To Desktop Version